What is the difference between PH-303, PH-302 and PHS-3301?
PH-303, PH-302 and PHS-3301 all support documented pH/ORP measurement, but their control interfaces and panel formats differ. Choose PH-303 for one isolated 4–20mA output (500 Ω max; ±0.05 mA accuracy), PH-302 for one 4–20mA output with RS485 Modbus RTU and two relays, or PHS-3301 for dual outputs, Modbus, more relay functions, data logging and a 144-series enclosure.
Which model should be selected first?
Start with the plant I/O list and the existing panel opening. A controller that measures the right parameter can still be the wrong choice if it lacks a required analogue channel, Modbus connection, relay function, recording feature or physical fit.
How do the three models compare?
| Model | Measurement and range | Outputs and communications | Relays and records | Format and limits |
|---|---|---|---|---|
| PH-303 | pH −2 to +16; ORP −2000 to +2000mV; temperature input | One isolated 4–20mA output (500 Ω max; ±0.05 mA accuracy); RS485 is not documented | Relay functions not documented on the current product source | 96-series; −10 to +130°C; confirm panel opening and output mapping |
| PH-302 | pH −2 to +16; ORP −2000 to +2000mV; temperature input | One isolated 4–20mA output, max. 500Ω; RS485 Modbus RTU; 9600/19200/38400 baud options | Two relay control points | 96-series; −10 to +130°C; 92.5 × 92.5mm cutout |
| PHS-3301 | pH −2 to +16; ORP −2000 to +2000mV; temperature input | Two isolated 4–20mA outputs, max. 500Ω; RS485 Modbus RTU; baud options | Two measurement relays plus one multifunction relay; up to 500,000 records | 144-series; −10 to +150°C; 141 × 141mm cutout |
The headline pH and ORP ranges are similar, so the practical selection usually comes from the signal path, relay logic, data requirement, temperature limit and panel size.
Is PH-303 suitable when Modbus is required?
Not from the current documented product information. PH-303 is documented with one isolated 4–20mA output, while RS485 Modbus RTU is documented for PH-302 and PHS-3301. If the PLC or DCS must read registers, review one of those two models and confirm the register map before quotation.
Is PH-302 suitable when two analogue channels are mandatory?
Not as the standard documented fit. PH-302 lists one isolated 4–20mA output, plus RS485 and two relay control points. If two analogue channels are mandatory, compare PHS-3301 and confirm its configuration, then confirm how pH/ORP and temperature are assigned to the outputs.
When is PHS-3301 worth considering?
PHS-3301 is worth reviewing when the project needs two analogue outputs together with RS485, multiple relay roles or local records. Its 144-series panel format and 141 × 141mm cutout must be checked before it replaces a compact 96-series controller.
What is each model suitable for?
PH-303 is suitable for
- projects requiring one isolated 4–20mA output;
- pH/ORP and temperature monitoring where the analogue receiving system is already defined;
- panel designs that can accommodate the documented 96-series format after confirmation.
PH-302 is suitable for
- compact 96-series panels needing one analogue channel, RS485 Modbus RTU and two relay control points;
- PLC/DCS integration where address, baud rate, register meaning and scaling can be commissioned;
- projects that do not need two analogue outputs or the PHS-3301 record capacity.
PHS-3301 is suitable for
- 144-series panels needing dual isolated 4–20mA outputs and RS485;
- applications requiring two measurement relays plus a multifunction relay;
- projects where up to 500,000 local records are relevant and the larger panel opening is acceptable.
What is each model not suitable for?
- PH-303: do not select it when RS485 Modbus or relay control is a mandatory documented requirement unless Tenber confirms a specific configuration.
- PH-302: do not select it when two analogue outputs, local record storage or a 144-series panel is required.
- PHS-3301: do not select it when the enclosure cannot accept a 141 × 141mm opening or the project does not need its larger feature set.
What must be confirmed before an inquiry?
Send the measurement parameter, liquid composition, expected pH/ORP range, temperature and pressure, installation point, electrode, temperature input, analogue channel count, Modbus requirement, relay logic, panel cutout, quantity and destination country through Tenber’s controller-selection inquiry page.
For the measuring loop, also review the industrial wastewater pH electrode selection guide and the pH/ORP controller range.
Frequently asked questions
Which model has RS485 Modbus RTU?
PH-302 and PHS-3301 document RS485 Modbus RTU. Confirm the required register map, address, baud rate, parity and commissioning responsibilities before integrating with a PLC or DCS.
Which model documents two isolated 4–20mA outputs?
PHS-3301 documents two isolated 4–20mA outputs; PH-303 has one isolated output. The better choice depends on whether RS485, relay functions, data logging and the 144-series panel format are also required.
Are the three controllers drop-in replacements?
No. Their documented output counts, communication features, relay functions and panel cutouts differ. Treat the controller, electrode, temperature input and enclosure as one configuration and confirm the wiring before ordering.
Sources and editorial record
- Current product sources: PH-303, PH-302 and PHS-3301.
- Current manual sources: PH-303 English operating manual (filename label: 2025 Replica; no revision date or number is printed in the PDF); PH-302 English manual; PHS-3301 English manual; and the public manual hub.
- Source review date: 2026-10-20. Author: Tenber Technical Content Team. Last updated: 2026-10-20.